Página 1 - WEIGHT INDICATING INSTRUMENT

210-FE WEIGHT INDICATING INSTRUMENT INSTALLATION and TECHNICAL MANUAL 8200-M242-O1 Rev B PO BOX 151 y WEBB CITY, M

Página 2 - 8200-M242-O1 y 210-FE

8200-M242-O1 y 210-FE Page 6 INSTALLATION Before beginning installation of your 210-FE Weight Indicating Instrument, make certain that the instrument

Página 3 - PRECAUTIONS

8200-M242-O1 y 210-FE Page 7 CAUTION! Disconnect any external load cell power supply before connecting load cells to the instrument. Failure to do s

Página 4 - PROPER DISPOSAL

8200-M242-O1 y 210-FE Page 8 INSTALLATION, Cont. 8. To terminate a wire, use a small flat blade screwdriver and press down on the release bar for the

Página 5 - SPECIFICATIONS

8200-M242-O1 y 210-FE Page 9 INSTALLATION, Cont. Load Cell Connections with Over 30 Feet of Cable For installations with over 30 feet of cable between

Página 6 - SPECIFICATIONS, Cont

8200-M242-O1 y 210-FE Page 10 INSTALLATION, Cont. Optically Isolated Inputs (requires additional hardware) Included with the I/O are 4 programmable in

Página 7

8200-M242-O1 y 210-FE Page 11 INSTALLATION, Cont. Relay Board - (Optional) The relay board (Cardinal p/n 8539-C062-0A) is mounted in the RB4-F externa

Página 8

8200-M242-O1 y 210-FE Page 12 INSTALLATION, Cont. Main PCB Figure No. 9

Página 9 - PRECAUTIONS, Cont

8200-M242-O1 y 210-FE Page 13 INSTALLATION, Cont. Main PCB Jumpers J1 - BATTERY CHARGE MODE (Not Applicable) J2 - AUTO-ON JUMPER The AUTO-ON jumper J

Página 10 - INSTALLATION

8200-M242-O1 y 210-FE Page 14 KEYPAD FUNCTIONS The Model 210-FE is equipped with a 22-key keypad. The keypad is used to enter commands and data into t

Página 11 - INSTALLATION, Cont

8200-M242-O1 y 210-FE Page 15 KEYPAD FUNCTIONS, Cont. 7. If the date displayed is correct, press the ENTER key to proceed to the consecutive number p

Página 12

8200-M242-O1 y 210-FE Page 2

Página 13

8200-M242-O1 y 210-FE Page 16 KEYPAD FUNCTIONS, Cont. PRINT KEY Pressing this key will add the displayed gross or net weight or piece count to the ass

Página 14

8200-M242-O1 y 210-FE Page 17 KEYPAD FUNCTIONS, Cont. NET/GROSS KEY This key is used to toggle between the Net and Gross weight modes. The selected m

Página 15

8200-M242-O1 y 210-FE Page 18 KEYPAD FUNCTIONS, Cont. 4 and NET/GROSS KEY, NET/GROSS KEY, ZERO KEY This combination will zero (clear) the Gross accumu

Página 16

8200-M242-O1 y 210-FE Page 19 ANNUNCIATOR DISPLAY The annunciator display is a three (3) character 16-segment display to the right of the weight displ

Página 17

8200-M242-O1 y 210-FE Page 20 ANNUNCIATOR DISPLAY, Cont. NOTE! The &=, 02, 3& or 7/ is alternately displayed or replaces the weight uni

Página 18 - KEYPAD FUNCTIONS

8200-M242-O1 y 210-FE Page 21 SETUP AND CALIBRATION Your Model 210-FE indicator has been thoroughly tested and calibrated before being shipped to you.

Página 19 - KEYPAD FUNCTIONS, Cont

8200-M242-O1 y 210-FE Page 22 SETUP AND CALIBRATION, Cont. DO NOT operate the keypad with pointed objects (pencils, pens, etc). Damage to keypad resu

Página 20

8200-M242-O1 y 210-FE Page 23 SETUP AND CALIBRATION, Cont. %++ (Password Y/N) The %++ (Password Y/N) prompt determines whether a password is require

Página 21

8200-M242-O1 y 210-FE Page 24 SETUP AND CALIBRATION, Cont. (1.<? (Weighing Unit 1) Press the ENTER key to show the current value. If the setting

Página 22 - DISPLAY BRIGHTNESS

8200-M242-O1 y 210-FE Page 25 SETUP AND CALIBRATION, Cont. <:"? (4% Zero Range) Press the ENTER key to show the current value. If the setting

Página 23

8200-M242-O1 y 210-FE Page 3 TABLE OF CONTENTS SPECIFICATIONS - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

Página 24

8200-M242-O1 y 210-FE Page 26 SETUP AND CALIBRATION, Cont. %A3"? (Preset Print-On-Balance) With only one Preset selected (4a$(<? 1 or 11), an

Página 25 - SETUP AND CALIBRATION

8200-M242-O1 y 210-FE Page 27 SETUP AND CALIBRATION, Cont. "<:? (Clear Tare) The Clear Tare feature allows the indicator to clear the Stored

Página 26 - SETUP AND CALIBRATION, Cont

8200-M242-O1 y 210-FE Page 28 SETUP AND CALIBRATION, Cont. A4 (A4) - Analog to Digital Filtering 4"<? (Digital Filtering) With A4 (A4)

Página 27

8200-M242-O1 y 210-FE Page 29 SETUP AND CALIBRATION, Cont. FILTER SETTING RECOMMENDATIONS Non Critical Sample Rate If the sample rate is not critical

Página 28

8200-M242-O1 y 210-FE Page 30 SETUP AND CALIBRATION, Cont. " (") - Calibration With " (") displayed, press the ENTER key

Página 29

8200-M242-O1 y 210-FE Page 31 SETUP AND CALIBRATION, Cont. Dual-Point with Zero (First Zero) Calibration "? – FIRST CALIBRATION WEIGHT The dis

Página 30

8200-M242-O1 y 210-FE Page 32 SETUP AND CALIBRATION, Cont. Single-Point for Span Only (Last Zero) Calibration "? – FIRST CALIBRATION WEIGHT Th

Página 31

8200-M242-O1 y 210-FE Page 33 SETUP AND CALIBRATION, Cont +.7 (+.7) - Serial Input/Output With +.7 (+.7) displayed, press the ENTER key. The displa

Página 32

8200-M242-O1 y 210-FE Page 34 SETUP AND CALIBRATION, Cont If SMA is selected, the data will be transmitted in the following format: <lf><s&g

Página 33

8200-M242-O1 y 210-FE Page 35 SETUP AND CALIBRATION, Cont If WI110 is selected, the data will be transmitted in the following format: <m><sp

Página 34 - CALIBRATION MODES

8200-M242-O1 y 210-FE Page 4 FCC COMPLIANCE STATEMENT WARNING! This equipment generates uses and can radiate radio frequency and if not installed and

Página 35

8200-M242-O1 y 210-FE Page 36 SETUP AND CALIBRATION, Cont %:.1< (%:.1<) - Print Tab Settings With %:.1< (%:.1<) displayed, press the ENT

Página 36

8200-M242-O1 y 210-FE Page 37 SETUP AND CALIBRATION, Cont 1<? (Net Weight Print Location) Press the ENTER key to show the current setting for the

Página 37

8200-M242-O1 y 210-FE Page 38 SETUP AND CALIBRATION, Cont +%1 (+%1) - Fine Span Adjustment NOTE! The +%1 and .a:+ modes require a load of 1

Página 38

8200-M242-O1 y 210-FE Page 39 SETUP REVIEW The 210-FE indicator allows several operational parameters to be reviewed and changed without breaking the

Página 39

8200-M242-O1 y 210-FE Page 40 CALIBRATION “C” NUMBERS The “C” numbers are displayed only during the Test mode operation by pressing the 4 key and then

Página 40

8200-M242-O1 y 210-FE Page 41 TROUBLESHOOTING Error Codes The 210-FE indicator is equipped with software that indicates when an error in the operation

Página 41

8200-M242-O1 y 210-FE Page 42 TROUBLESHOOTING, Cont. Error Codes, Cont. = aUNITS key pressed in an attempt to perform a “unit” conversion that is no

Página 42

8200-M242-O1 y 210-FE Page 43 TROUBLESHOOTING, Cont. Before You Call Service The 210-FE indicator has been designed to provide you with years of troub

Página 43 - SETUP REVIEW

8200-M242-O1 y 210-FE Page 44 CALIBRATION SEAL INSTALLATION If your 210-FE Weight Indicating Instrument is used in a commercial application it must be

Página 44 - ACCUMULATORS

8200-M242-O1 y 210-FE Page 45 AXLE WEIGHER MODE OPERATION After placing the wheel load weighers on the ground in front of the vehicle(s) to be weighe

Página 45 - TROUBLESHOOTING

8200-M242-O1 y 210-FE Page 1 SPECIFICATIONS Power Requirements: 90 to 264 VAC (50/60 Hz) at 0.4A Enclosure Type: NEMA 4X/IP66 Enclosure Size: 9.

Página 46 - TROUBLESHOOTING, Cont

8200-M242-O1 y 210-FE Page 46 AXLE WEIGHER MODE OPERATION, Cont. Additional Features If you wish to also record the total of any axle groups, press th

Página 47

8200-M242-O1 y 210-FE Page 47 PART IDENTIFICATION (Final Assembly) ITEM NO. QTY. PART NUMBER DESCRIPTION 1 1 8200-D296-0A SUB ASSEMBLY: FRONT E

Página 48

8200-M242-O1 y 210-FE Page 48 PART IDENTIFICATION, Cont (Front Enclosure Sub Assembly) ITEM NO. QTY. PART NUMBER DESCRIPTION 1 28 6013-0039 NUT

Página 49 - AXLE WEIGHER MODE OPERATION

8200-M242-O1 y 210-FE Page 49 PART IDENTIFICATION, Cont (Front Enclosure Sub Assembly)

Página 50 - Additional Features

8200-M242-O1 y 210-FE Page 50 PART IDENTIFICATION, Cont (Rear Enclosure Sub Assembly) ITEM NO. QTY. PART NUMBER DESCRIPTION 1 5 6013-0315 NUT #

Página 51 - PART IDENTIFICATION

8200-M242-O1 y 210-FE Page 51 PART IDENTIFICATION, Cont (Rear Enclosure Sub Assembly)

Página 52 - PART IDENTIFICATION, Cont

STATEMENT OF LIMITED WARRANTY WARRANTY TERMS Cardinal Scale Manufacturing Company warrants the equipment we manufacture against defects in material a

Página 53

PRODUCT TYPE TERM MATERIAL AND WORKMANSHIP LIGHTNING DAMAGE See note 9 WATER DAMAGE See note 7 CORROSION See note 4 ON-SITE LABOR LIMITATIONS AND REQU

Página 54

APPLICABLE LIMITATIONS AND REQUIREMENTS 1. This warranty applies only to the original purchaser. The warranty does not apply to equipment that has b

Página 55

EXCLUSIONS This warranty does not include replacement of consumable of expendable parts. The warranty does not apply to any item that has been damage

Página 56 - See note 9

8200-M242-O1 y 210-FE Page 2 SPECIFICATIONS, Cont. Standard Features: • Push button tare function • Gross, tare, net conversion • Selectable key lo

Página 57 - 315-WARRANTY-CAR-E

8200-M242-O1 y 210-FE Page 56

Página 58

8200-M242-O1 y 210-FE Page 3 EUROPEAN DECLARATION OF CONFORMITY Manufacturer: Cardinal Scale Manufacturing Company PO Box 151 203 Eas

Página 59

8200-M242-O1 y 210-FE Page 4 PRECAUTIONS Static Electricity CAUTION! This device contains static sensitive circuit cards and components. Improper h

Página 60 - Page 56

8200-M242-O1 y 210-FE Page 5 PRECAUTIONS, Cont. Care and Cleaning 1. DO NOT submerge the indicator in water, pour or spray water directly on it. 2. DO